Replacement Procedures
8-18 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
8-3-6-5 Cleaning the metal filter
Clean the system's metal filter to ensure that a clogged filter
does not cause the system to overheat and reduce system
performance and reliability. It is recommended the filters be
cleaned every two weeks, but the requirements will vary due to
your system use.
1. Pull out the air filter.
Figure 8-7. Pull Out the Air Filter
CAUTION
Be sure to lock the wheels before cleaning the metal filter to
avoid injury by any unexpected movement of the system.
DO NOT operate the unit without the metal filter in place.
Allow the metal filter to dry thoroughly before re-installing them
on the unit.