Replacement Procedures
8-26 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
8-4-7-1 Loading the System Software with USB memory stick(continued)
NOTE: If selected to load C Disk only when loading the system
software, the system will display a screen to restore
Computer name before entering the scanning screen.
9. If Load The Complete Disk is selected, a confirmation
page is displayed, then select Execute to continue.
Figure 8-14. Data Backup Confirmation
10. System USB memory stick will be loading. Wait for the
software installation to complete. Status bar on the screen
indicates the progress.
NOTE: It will need about 27 minutes to finish the upgrading
process.
While the software install procedure is designed to
preserve data, you should select choice [b] to format disk
C only.
Please make sure the system configuration have been
backed up first, then upgrade the system.
Do not interrupt the software loading at any time..