Replacement Procedures
8-28 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
8-4-7-1 Loading the System Software with USB memory stick(continued)
12. When the system reboots, the Installation Wizard page
shows. You can configure the system settings or exit. For
detailed process, please refer to 8-4-13 ‘Installation Wizard’
on page 8-34.
13. If Upgrade the Whole Disk is selected, you are required to
input the serial number when you reboot the system. After
you press OK, the system will shut down automatically.
Figure 8-17. Input Serial Number
14. After you power on the system, it will need you to input the
backup option key, and then press OK.
Figure 8-18. Input Option Key