Safety considerations
Vivid T9/Vivid T8 – Basic Service Manual 1-15
5795591-100 English Rev. 3
1-6-3 Human Safety(continued)
WARNING
Risk of electrical shock, Ultrasound system must be turned off
and disconnected from power source. Cord must be controlled
at all times.
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The white light on the
OP panel on/off button will turn off.
Ultrasound System components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions
WARNING
Use all Personal Protection Equipment (PPE) such as gloves,
safety shoes, safety glasses, and kneeling pad, to reduce the
risk of injury.
WARNING
Beware of possible sharp edges on all mechanical parts. If
sharp edges are encountered, the appropriate PPE should be
used to reduce the risk of injury.
WARNING
Wear all PPE including gloves as indicated in the chemical
MSDS.