GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
3-50 Section 3-7 - Printer Installation
3-7-7-1 UP-D897 / UP-D898MD - Printer Settings
1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-7 "Adjustment of Printer Settings" .
2.) Select the
LAYOUT page and select:
- Paper: 960x1280
- Orientation: Portrait
- Interpolation Method: Bilinear
3.) Select the
DENSITY ADJUST page and select:
- Gamma: TONE2
- Advanced = 0; Dark = 0; Light = 0 ; Sharpness = 2
4.) Save the adjusted printer settings click APPLY
and then OK.
Close window and exit System Setup with SAVE&EXIT
.
5.) Assign the Printer to the remote keys P1
, P2, and/or P3 with the save button;
see: Section 3-7-8 "Remote Control Selection" on page 3-53.
Figure 3-50 Layout + Density Adjust page