GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-179
Section 8-58
Replacement of Battery Pack Assy
8-58-1 Manpower
One person, 20 minutes
8-58-2 Tools
Philips screwdriver
8-58-3 Preparation
Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-58-4 Removal procedure
1.) Follow the removal procedure of Section 8-41 "Replacement of the DPS/DPS with Battery" on page
8-128.
2.) Unscrew 14 screws and remove DPS with Battery cover.
Figure 8-257 Removing DPS with Battery cover