GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-85
7.) Select the CONNECT button to establish the connection to the remote machine.
If successful, the DISCONNECT button becomes active.
NOTICE
!! NOTICE:
If there is an error during the connection, a warning message appears inside the dialog.
In this case, please verify the data in the dialog.
NOTICE
!! NOTICE:
If there already is a connection to the remote server, the CONNECT button is grayed.
To change the existing connection, first click on DISCONNECT
and then enter the new settings.
WARNINGWARNING
!! WARNING:
Please make sure that the server you are connecting to is trustworthy and reliable.
For details, contact your local system administrator.
If you backup Archive data to this server, all the patients’ demographic data will be
copied to this server!
NOTICE
The Map network Drive button is also accessible in the “Connect USB and Network Drives” dialog
window that appears when pressing the Eject key on the alphanumeric keyboard.