GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-31
1.) If not already connected, plug all probes to be tested to the Voluson S&P - series system.
2.) Check mark ProbeAssessmentTool in CSD - Diagnostics - HW Tests.
Probe diagnostic tests will run for all connected probes and show up as completed after few
minutes.
3.) Move the scroll bar downwards to view specific probe assessment results of all connected probes.
Figure 7-31 ePAT - Test results