GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-39
Gather Logs Utility
This will gather up logs and presets. Logs are zipped up and located in (D:\export\Logs_xxxx.zip) for
retrieval by the OnLine Center.
Image Viewer Utility
A list of all images stored in D:\export is displayed. It is possible to display the Images.
Image Compress & Delete Utility
A list of all images stored in D:\export is displayed. It is possible to compress or delete the select the
images. The compressed files are added to (D:\export\xxxx.zip).
Virtual Console Observation
This offers GE technicians the possibility to view the entire customer´s desktop and operation system.
Using VCO (Virtual Console Observation) a service technician or OLC can access and modify all PC
settings and programs or run diagnostics on the customer´s ultrasound system
Disruptive Mode can be selected by the customer directly on the Voluson S&P-Series (see How the
Customer enables/disables Disruptive Mode and VCO), or requested remotely by the service technician
or OLC (see How the FE remotely enables/disables Disruptive Mode and VCO ).
NOTE: ONLY registered GE Service personnel have access to this feature. A password is required. All actions
and changes should be documented.
Invoke Service Agent FastPoll
Invokes FastPoll (fast polling) mode.