124 XL Vu™ VideoProbe
®
Appendix
B. Chemical Compatibility
Chemical compatibility refers to the probe’s ability to come into contact
with various liquid substances and not be damaged.
WARNING Do not use this system in explosive environments.
These substances are safe for a short duration of contact with the
insertion tube:
Water
Aircraft Gasoline
Jet-A-Fuel
Isopropyl alcohol
JP-4 Fuel
Kerosene
Synthetic Turbo Oil
Gasoline
Diesel Fuel
Hydraulic Oil
Inhibited Transformer Oil
Subsequent to contact with the above uids, the insertion tube must be
cleaned prior to storage.