Mechanical Separation / GEA Westfalia Separator
4.15 Alarm #21 - WMS sensor fault (option)
WMS – SENSOR FAULT
During WMS CHECK
During TIMER 45
Alarm#21 ENTER
Text display Possible causes Action
WMS sensor soiled.
•
Check WMS sensor.
WMS sensor malfunction.
•
Check WMS sensor.
Water discharge valve V6 doesn’t
open.
• Check function V6.
Not sufficient water during T45
• Check water volume.
T45 too short.
•
Check T45.
T02 too short (HFO).
•
Check T02.
•
Deselect function if required.
Help text – information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
WMS – SENSOR FAULT
Please check:
WMS sensor
soiled
->
Page 2
Function WMS
Sensor
Function V5/V6
Water volume
->
Page 3
Timer 45
Timer 02 (HFO)
too short
back with ENTER
F1