4.20 Alarm #33 - Ejection monitoring (option)
Eject. speed < offset
Separation “Stop”
Alarm#33 ENTER
Text display Possible causes Action
Operating water valve V3 does not
open.
• Check bowl hydraulic function with
Connector on valve V3 loose.
•
Offset for ejection speed adjusted too
high.
• Check offset setting.
Water volume at V3 too low.
•
Monitoring time T28 set too short.
•
Speed pickup defective.
• Check speed pickup electrically;
Filter insert in strainer at solenoid
block soiled.
• Check filter insert.
Clutch shoes worn or oily.
Note :
Alarm #33 only active when speed pickup is installed!
Help text – information on the control system
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
EJECTION MONITORING
Please check:
Function bowl
hydraulics with V3
Page 2
Connector valve V3
Speed< offset
Water volume
T28
Page 3
Speed pickup
electric
Pre-strainer
back with ENTER