11.3 Service
In case the VisualiX system does not take images it is possible to check whether the
sensor or the IME is not functioning part. To do that simply right click on the Status
Icon and select “Service” to open the Service tab. In this tab the user can also see the
last obtained image. By pressing the “NoXrayGrab” button a sample image should be
generated when the IME works properly. Note that no exposure to X-rays is necessary
for the creation of this image.
11.4 Components Code List for VisualiX
Description Code
1 VisualiX eHD IME and X-ray image sensor (Universal size) 4519 105 03601
2 VisualiX eHD IME and X-ray image sensor (Large area) 4519 105 03611
3 Intraoral aiming device 9868 998 00011
4 Disposable protection bags for Universal Size sensor 9869 999 00011
5 Disposable protection bags for Large Area sensor 9869 999 00011
6 USB cable 4.5 metres long 4519 105 03201
7 USB cable 2 metres long 4519 105 03401
8 Accessories
• Wall mount IME adapter
• Sensor hook with adesive (ts both sizes)
4519 105 03591
4519 105 03581