44-00-0297 SHERLOCK102/202 REV. 4.1 05-02-15
29
This menu allows the user to veiw but not change the various
controller parameters. Screen entries in “<STATUS MENU>”
will only be shown if their respective control options are
installed in the “<SYSTEM CONFIG MENU>”
At DEFAULT SCREEN, press UP/DOWN button to scroll to
the "<STATUS MENU>". Press the SELECT button to enter
this menu. "<STATUS MENU>" is shown on the first line of
the display. The control status is shown on the second line of
the display. These control status are:
The first screen shows the real time clock.
Oxygen Sensor Current reading if installed at the snesor A
input.
Oxygen Sensor Alarm Setpoint.
Sensor A current reading if a refrigerant sensor was installed in
the sensor A input..
Sensor A level 1 alarm setpoint
Sensor A level 1 alarm delay in minutes
Sensor A level 2 alarm setpoint
Sensor A level 2 alarm delay in minutes
Sensor 1 level 1 setback setpoint
Sensor 1 level 2 setback setpoint
Continued on next page
<STATUS MENU>
SELECT/
CHANGE
<STS> 16:30
05/02/08
STATUS MENU
REPEAT FOR SENSOR B
<STS>LEAK SNR A
READING: 0 PPM
<ST> SNR A LEVEL1
ALARM: 150 PPM
<ST>SNR A LEVEL1
ALM DELAY: 2M
<ST>SNR A LEVEL2
ALARM: 250 PPM
<ST>SNR A LEVEL2
ALM DELAY: 2M
<STS>SNR A SETBK
LVL1 ALM: 150 PPM
<STS>SNR A SETBK
LVL2 ALM: 250 PPM
REPEAT FOR SENSOR B
<STS>OXYGEN SNR
READING: 20.9 %
<ST>OXYGEN LEVEL
LOW ALARM: 17.0 %