4. Operation
4.1 Key Functions of Control Panel
1. RPM/RCF
RPM/RCF indicates rotating speed.
The maximum RPM/RCF of 1730R is 17,000/27,237.
2. TIME
Time can be set at ‘Minute’ and ‘Second’ with the range up to 99 min 59 sec (00:00:
continuous).
3. TEMP
Temperature can be set from -20 degree to 40 degree. When [TEMP] button is kept pressed for
longer than 2 seconds, it can reach rapid refrigeration up to the setting temperature value (Fast
Cool).
4. ACC/DEC
The acceleration and deceleration rates can be set from1 to 9 and from 0 to 9, respectively. The
DEC ‘0’ in deceleration means natural deceleration. As the number increases,the acceleration
or the deceleration becomes faster.
5. PROG (Program)
Max. 100 programs can be saved and recalled.
When [PROG] button is kept pressed for longer than 2 seconds, all buttons can be locked or
unlocked.
6. UP/DOWN
When setting up the RPM/RCF and Time, you can put the set value with up and down button.
7. ENTER
[ENTER] button is for completion of data setting.
8. PULSE
[PULSE] button is for quick and short spin-down. Rotor spins whilethe button is kept pressed
and stops immediately as the button is released.