Refrigerated High-Speed Micro Centrifuge, 1730R
17
4.7 PULSE (Short Spin)
This function can be used for the quick and short spin-down. If the touched pulse [] button is
released, the centrifuge decelerates immediately. (Short Spin)
Action
4.8 Save / Call Program
Action
Touch [ ]button and release at the point you want to stop the
operation. And then, the centrifuge decelerates immediately.
If the actual speed reaches the set value, the centrifuge starts to
decelerateeven when [PULSE] button is kept pressed.
Set parameters (RPM/RCF, Time, Temperature and ACC/DEC,
etc.) and savethem in the program. (refer to 4.2 Setting Rotation
Speed/ 4.3 Setting Time/
4.4 Setting Temperature and Fast Cool / 4.5
Setting Acceleration / Deceleration)
Touch [PROG] button twice.
The program number ickers and the SAVEsign appears
on the display window.
Touch Up[ ] or Down [ ] button to select the program
number and touch [ENTER] button to complete the saving.
If you do not touch the Up/Down button for 15 seconds, the
setting mode is cleared.
Maximum 100 programs can be saved. (Program number: 00 to
99)
1
1
2
3
Save Program
The set parameters (speed, time and ACC/DEC) can be stored in the individual programs
and recalled for next use.
Call Program
The saved programs from 0 to 99 can be recalled at need.