Refrigerated High-Speed Micro Centrifuge, 1730R
3
This manual is for the users who operate the device for the rst time.
This manual provides information on the detailed instructions, precaution,
troubleshooting and maintenance care.
1. Meanings of Symbols & Safety Precautions
1.1 Safety Labels on Instrument
The labels on the instrument represent safety instructions and directions.
Symbol Meaning Symbol Meaning
Attention and warning
Attention and warning for
electric shock
Symbol Meaning
Attention and warning for correct way of
samplebalancing in the rotor.
Attention and warning for rotor coupling.
Attention and warning for door opening and closing.
1.2 Safety Precautions
Before using the instrument, please read this operation manual to ensure correct usage. Incorrect
handling of the instrument could possibly result in personal injury or physical damage on the
instrument or its accessories.
1. ALWAYS locate the instrument on a at, rigid and stable table capable of withstanding the weight
of the instrument and its spinning operation.
2. ALWAYS make a
safety zone of 30 cm around the centrifuge to indicate that neither hazardous
materials nor persons should be permitted within the area during operation.
ALWAYS position the instrument with enough space on each side of instrument to ensure
proper air circulation.
3. ALWAYS install the instrument within a temperature and humidity controlled environment.
(Permissible ambient temperature: +5 ~
+35 , Relative humidity: 85%)
4. Before connecting the power, check the rated voltage.
5. Should not use unapproved rotors and associated accessories.
Only use rotors from Gene Company Limited with appropriate centrifugal tubes and suitable
adaptors to embrace sample containers tightly enough inside rotors.
6. Before operating the instrument, check if the rotor and the rotor lid are secur
elyfastened.
Should operate the instrument with a rotor properly installed and secured to the motor shaft.