4.2 Setting Rotation Speed (RPM/RCF)
The rotation speed is indicatedas RPM (Rotation per Minute)/ RCF (Relative Centrifugal
Force).
The speed range is 500~17,000 rpm and 24~27,237 xg.
Action
Touch [RPM/RCF] once.
The RPM mode is activated bytouching [RPM/RCF].
If the RCF mode is selected, touch [RPM/RCF] once more.
The input RPM value flickerson the display window and can be
set here.
TouchUp [ ] or Down [ ] button to change input value.
RPM setting unit: 10 rpm
RCF setting unit:1xg
If you keep holding nger on the Up/Down buttons for
more
than 5 seconds, the RPM/ RCF value is gradually increased or
decreased with 100 rpm, 1000 rpm unit and 10 xg, 100 xg and 1000
xg unit.
If you do not touchthe Up/Down button for 15 seconds, the
setting mode is cleared.
Touch the [ENTER] button to save the setting value.
1
2
3