Refrigerated High-Speed Micro Centrifuge, 1730R
15
4.5 Setting Acceleration/ Deceleration (ACC/DEC)
Acceleration (level: 1~9) and deceleration (level: 0~9 (0: spontaneous deceleration)) can be
controlled according to the sample characteristics
Action
Touch [ACC/DEC] button.
The input value of the ACC or DEC valueappears on TEMP
position of the display window and can be set here.
Touch Up [ ] or Down [ ] button to change input ACC or
DEC value.
Input the desired level of ACC from 1 to 9. (Level 9: The
fastest acceleration)
Input the desired level of DEC from 0 to 9. (Level 0: Natural
deceleration / Level 9: The fastest deceleration).
If you do not touch the Up/Down button for 15 seconds, the
setting mode is cleared.
Touch [ENTER] button to complete the setting.