5. Maintenance
5.1 Outer Part of Instrument, Chamber, Shaft and Rotor
Outer Part of Instrument
1. Clean the outside of the instrument with dry soft cloth. If necessary, dip the cloth in neutral
detergent and clean any contaminated area. Keep completely dry after cleaning.
2. Do not use any volatile chemicals such as alcohol, benzene, benzole, and thinner, etc.
3. Be careful not to make scratches on the surface of the instrument.
Scratches may cause corrosion on the surface of the inst
rument.
Any parts with rust should be cleaned with neutral detergents and kept dry.
Chamber
1. Keep dry inside the chamber after every use.
2. If the chamber is contaminated, clean contaminated area with the cloth dipped in neutral
detergent.
Shaft
1. Always keep the motor shaft clean to avoid any imbalance problem caused by the
contaminants.
2. After using the instrument, take out the rotor from the shaft a
nd clean the shaft with dry soft
cloth to keep dry.
3. If the rotor cannot be easily removed from the shaft, do not pull the rotor by force and call a
service engineer authorized by Gene Company Limited.
Rotor
1. If any parts become contaminated, clean them with soft wet cloth and keep the rotor dry.
2. Be careful not to make scratches inside or on the surface of rotors. Any small scratches can
cause corr
osion of the rotor and big damage to the instrument.
3. While the instrument is not used, remove the rotor from the motor shaft and stand it upside
down.