42
C. Gerhardt GmbH & Co. KG
WAIT FOR STEAM
Message disappears as soon as the steam ge-
nerator has reached the necessary pressure.
Start analysis by pressing "Enter".
CHECK CHEMICAL RESERVOIRS
If using set of tanks KAN 40 check tanks.
If using other tanks, switch off "Chemical Reser-
voirs" (see chapter 7.6. "Chemical Reservoirs").
The analysis is continued.
Error message disappears only after error is cor-
rected.
Error message Actions
CORRECT TITRE/FACTOR
THE FOLLOWING PROGRAMS HAVE
NOT BEEN DEFINED
THE FOLLOWING PROGRAMS HAVE
BEEN INCORRECTLY DEFINED
The defined values for the series in process have
to be kept. They can be called up in the main
menu
"Statistics" (see chap. 12 "Statistics").
If a series is called up which uses a program that
has been deleted after sample input the program
has to be defined again. Call main menu "Pro-
gramming" and check programming.
If a series is called up that uses a program in
which the units (weights, volumes) have been
changed the program has to be defined again.
Call main menu "Programming" and check pro-
gramming.
NO COOLING WATER
Cooling water pressure < 0,5 bar. Check cooling
water inlet. Make sure tap is open completely.
Press "Enter" to continue analysis.
Press "Stop" to interrupt analysis.
RECEIVER HAS OVERFILLED
Analysis is stopped.
Empty the receiver (take out glass or start suc-
tion
manually).
NO SAMPLE TUBE
Insert tube and restart by pressing "Enter".
If this happens during an analysis it is interrup-
ted.
ERROR MESSAGES