13. Error messages
All functions of the distillation system are controlled continuously. As soon as there is an error it is
indicated on the display as well as by an acoustic signal.
In most cases the error message on the display can be erased by gently touching the display. But the
error as such is not corrected by doing so. This chapter will inform you what actions to take if there is
an error message.
13.1. General error messages
Error message Actions
SENSOR STEAM GENERATOR FAULT
Turn off equipment and turn on again. If failure
persists consult technical support.
OVERPRESSURE STEAM GENERATOR
Turn off equipment and turn on again after ap-
prox. 15 minutes. If failure persists consult tech-
nical support.
NO H
2
O FOR STEAM GENERATOR
Turn off equipment, check tank for H
2
O, turn on
again. If failure persists consult technical support.
ERROR MESSAGES
ATTENTION! CORRECT DATE
The battery might be damaged and therefore, the
date is overwritten.
Type current date. If failure persists consult tech-
nical support.
13.2. Error messages during the analysis
Error message Actions
ATTENTION!
95 % OF THE MEMORY IS TAKEN. 500
RESULTS CAN BE STORED
ANALYSIS CANNOT BE STARTED FOR
RESULTS. MEMORY IS TAKEN COM-
PLETELY
The results stored on the hard disk have to be
deleted in order to provide storage capacity for
future results (see chapter 12 "Statistics"). This
message has to be confirmed by pressing "Enter".
The system points out that the storage capacity
of the hard disk is nearly taken and only 500
more analysis are possible. By deleting results
already stored more storage capacity on the hard
disk is available (see chapter 12 "Statistics"). This
message has to be confirmed by pressing "Enter".