4.3 Starting program
WARNING!
Risk of contaminated items
Incorrect selection of a program can affect the
cleaning result.
- Follow operation instructions in User manual.
- Visually inspect the items after unloading.
- Follow manufacturer's reprocessing instructions
for medical items.
WARNING!
If the machine has not been used for more than
72 hours, a wash program with an empty chamber
must be run before using the machine to clean
goods.
WARNING!
A fine filter must be used to clean and disinfect
tubular instruments with a passage of less than 2
mm, to prevent particles from getting stuck in the
instruments. Blocked or reduced water flow may
result in ineffective cleaning and no disinfection.
WARNING!
If the program is aborted before it is completed,
the goods will not be properly washed or
disinfected. The program must then be run again.
4 OPERATION
20 (56) User Manual Getinge 46-series