5 DETERGENT
When handling detergents, follow the safety
instructions as indicated on the detergent
container label and/or in its material safety data
sheet.
When selecting detergents, read and follow the
instructions and recommendations of the
manufacturer of the goods to be cleaned or
disinfected.
The following table is a guide.
Always make decisions regarding what detergent to use for
cleaning or disinfection of a specific load (material, soil and type of
items) together with the detergent supplier.
Make sure the intended use for the detergent corresponds with the
program and load.
Always contact a service technician for information regarding what
detergent to use in a specific program phase.
Area of use Detergent type
Thermal disinfection of non-invasive, non-critical
medical devices, non-blooded waste, aids for the
disabled, footwear and wash bowls.
Alkaline
Enzymatic
Thermal disinfection of non-invasive, non-critical
medical devices, blooded waste, container
systems used to transport medical devices, aids
for the disabled, footwear and wash bowls.
Alkaline
Enzymatic
Thermal disinfection of invasive and critical
medical devices, such as invasive equipment,
soiled, rigid and non-cannulated surgical
instruments.
Alkaline
Enzymatic
Chemical disinfection of non-invasive, non-critical
medical devices, non-blooded waste, aids for the
disabled, footwear and wash bowls.
Alkaline
Enzymatic
Chemical disinfection of non-invasive, non-critical
medical devices, blooded waste, container
systems used to transport medical devices, aids
for the disabled, footwear and wash bowls.
Alkaline
Enzymatic
For neutralizing alkaline residues. Acidic
5 DETERGENT
24 (56) User Manual Getinge 46-series
<Doc_USR><Doc_6002267102><Rev.B><Lang_en>