Area of use Detergent type
For removal of rust, hard water scale and stains
on instruments and in washer-disinfector
chamber.
Acidic
For use in the final rinse to speed up the drying of
the load.
Rinse aid
For use in the final rinse to lubricate moving parts
of surgical instruments and prevent friction
corrosion.
Lubricant
5.1 Using detergent
WARNING!
Dosing monitoring only works with detergents
with viscosities lower than 10 cSt.
WARNING!
Do not use detergent for household dishwashers.
Note!
The specified performance may not be achieved if other
process chemicals than those that were tested during type
testing are used.
The machine has been proven and tested with Getinge Disinfection
AB's range of detergents as per ISO 15883-1:2006.
• If chemicals are recommended in the manual, this does not
mean that the manufacturer of the machine accepts liability
for any impact the agent has on goods to be washed.
• Changes in the composition, conditions, etc. that were not
reported by the manufacturer may affect the result of the
cleaning process.
• Always follow the instrument manufacturer's instructions
when you use new instruments.
• The dosage amount depends on the water quality. For more
information, see the process chemicals product
specifications.
5 DETERGENT