6.3.2 Service table
Note!
The service table is only shown for information purposes.
WARNING!
Maintenance may only be performed by
authorized service technicians.
Interval
Component Check
yearly
Replace
if
necessa
ry
Replace
yearly
Replace
every
second
year
Replace
after 1000
hours
Replace
ment
time
Door seal • • 1 hour
Hoses between dosing pump
and detergent container
• • 15 min
Hose between dosing pump
and machine
• • 20 min
Filter in incoming media • • 30 min
Hose in hose pump • • 10 min
Spray arm bearings • • 5 min
Sterile filter in dryer • • 30 min
Hoses to dryer • • 1 hour
Check valve in dryer • • 1 hour
Check valve in drainage • • 30 min
Fan to dryer • • 1 hour
Spring on door closing • 1 hour
Dampers on door opening • 1 hour
Check that the main circuit
power cables are firmly
attached to the connections.
• 10 min
Check that the suction lance for
detergent dosing is correctly
labelled.
•
6.4 Machines that are not in use
A machine that is not being used should be drained. Contact a
service technician.
6 MAINTENANCE
User Manual Getinge 46-series 39 (56)
<Doc_USR><Doc_6002267102><Rev.B><Lang_en>