24
Information about the last twenty faults is saved in the control system
and is accessible in the service menu. If the sterilizer has a printer, this
information can be printed out.
Service codes
When the word “SERVICE” appears on the display, service is required
and must be carried out by a service technician. See below for
explanations of the codes.
NOTE!
No error has occurred.
The sterilizer can still be used.
Error code/
Message
Explanation
Battery fault
(Message)
The backup battery of the control system is
flat. Programs and stored parameters may
have been lost.
Temp. sensor
error chamber
Chamber temperature sensor defective.
Temp. sensor
error jacket
Jacket temperature sensor defective.
Pressure sensor
fault
Pressure transducer defective.
Mistakes when correcting the above types of faults may
lead to situations where there is a risk of injury.
Always call a trained service technician.
Error code Explanation
Pressure fault The chamber pressure has been too high
during the sterile phase.
Jacket fault The jacket temperature has been too high
during part of the process.
Gasket fault Door gasket pressure too low.
Door fault Door(s) not closed and interlocked.