25
Pump fault Vacuum pump has stopped.
High temperature The chamber temperature has been above the
permitted limit during the sterilization phase.
Low temperature The chamber temperature has been below the
permitted limit during the sterilization phase.
Time fault phase The phase named on the display exceeded the
time limits for completion. Check whether the
cause is a component fault or media fault.
Time fault post-
treatment
The post-treatment exceeded the times limits
for completion. Check whether the cause is a
component fault or media fault.
Service
(Message)
This message appears when the programmed
service interval has elapsed and the current
process is complete. The message continues to
be displayed until it is de-activated by the
service technician.
The autoclave can still be used.
Call a service engineer immediately.
Leak test fault An automatic leak test has failed.
Supervisor error The Supervisor’s independent monitoring of
process time and temperature conditions has
indicated that these have not been fulfilled and
have triggered an alarm.
Power failure There has been a power failure longer than 10
seconds.
Emergency stop The program has been stopped with the
emergency stop button.
Steam generator
fault
The supply of feedwater to the steam
generator has stopped working.
Stop unloading
(Message)
An unloading cycle with an automatic feeder
or discharger has not been completed
correctly.
Door interlock The independent safety interlocking of media
admission to the chamber, as controlled by the
door switch, is faulty.
Gasket
interlocking
The independent safety interlocking of media
admission to the chamber, as controlled by the
gasket pressure switch, is faulty.
Error code Explanation