38
What to do if there is an alarm
Programs are set up to terminate automatically.
The program ends automatically.
A process or a program that was stopped by emergency stop, which
is not terminated automatically
Note the phase of the process where the fault occurred.
Options for action are described under “Alarms” in the Operation
chapter.
After a fault in a process, the password must be entered before the
loading side door can be opened.
On power failure or if the control switch or working switch
of the sterilizer had to be turned off during a process, there
is a risk of leakage at the door. Safety measure when there
is positive pressure in the chamber:
KEEP WELL AWAY FROM THE DOOR AND CALL A
TECHNICIAN!
ALWAYS LEAVE THE PREMISES and call a technician if
the chamber pressure is in the RED area of the pressure
gauge.
Silence the audible signal by pressing
one of the buttons.
/ [ALARM OK]