5 ELECTRICAL INSTALLATION
1
2
3
4
005675
1. Isolator switch (not
included)
2. Supply cable
(not included). To be
sized by end user
according to local
conditions and
regulations
3. Supply cable
(3 m included)
4. Machine type
plate
5.1 Electrical connections
5.1.1 Connect electricity
• Installation should only be carried out by authorized staff.
• Check the machine type plate.
• Install the machine in accordance with installation category
CAT II.
• Make sure the electrical connection is carried out in
accordance with current local regulations and rules.
5.1.2
Isolator switch
• The machine must be preceded by a lockable isolator switch
for the electric power supply.
• The switch must be easily accessible on a wall close to the
machine.
• The installation must conform to and be marked in
accordance with local regulations.