7 ERROR CODE 1
Error code 01 indicates that contact with the temperature gauge
has been broken.
7.1 Drying anaesthetic and respiratory hoses and bellows
The drying cabinet can be used for 6 cassettes at the same time.
To enable the drying air to be forced through the tubing, each
drying cabinet is also equipped with 4 loose blind flanges. If fewer
than 6 cassettes are used during the drying operation, these blind
flanges must be slid into the empty guides. If only a single cassette
is used, all 4 blind flanges must be slid in, leaving one guide
uncovered.
7.2 Maintenance when necessary
When necessary, the outside and the inside of the drying cabinet
can be cleaned and disinfected with an alcohol-based detergent,
surgical spirit, or normal detergent. The control panel can be
cleaned in the same way. Remove marks with ordinary stain-
remover for stainless steel.
7 ERROR CODE 1
USER MANUAL GETINGE DRYING CABINET S-363 15 (24)
<Doc_6001300802><Rev.C><Lang_en>