3 TECHNICAL DATA
Humidity Max 80 % at 31 °C
Room temperature 5–40 °C
Weight Approximately 100 kg
Internal dimensions 1600 x 495 x 375 (H x W x D)
Fan capacity 200 m³/h
Exhaust air Fully open exhaust valve:
approximately 75 m³/h, max 90 °C.
Fully closed exhaust valve:
approximately 40 m³/h, max 90 °C.
Exhaust stub pipe Ø 100 mm valve
regulated.
Heat dissipation into the
room
Approximately 500 W if the exhaust
stub pipe is connected to the
exhaust ventilation.
Electrical Max current
(A)
Power
(kW)
Max Pre
Fuse
200 V, 3+PE, 50/60 Hz 11.6 A 2.6 kW C16 A
230 V, 3+PE, 50/60 Hz 13.3 A 3.4 kW C16 A
240 V, 3+PE, 50/60 Hz 13.8 A 3.7 kW C16 A
400 V, 3N+PE, 50/60 Hz 7.3 A 3.4 kW C10 A
415 V, 3N+PE, 50/60 Hz 7.8 A 3.7 kW C10 A
Pollution degree 2
Max altitude 2000 m
Indoor use only
Overvoltage category 2
Main supply voltage
fluctuations
+- 10% at nominal voltage short
term.
3 TECHNICAL DATA
8 (24) USER MANUAL GETINGE DRYING CABINET S-363