How to resolve airway temperature high alarm on GGM Humidifier?
- PpacemonicaAug 27, 2025
If the airway temperature high alarm is activated on your GGM Humidifier, wait for the temperature to fall below the threshold if there was a sudden spurious increase in gas flow. If the heater wire is too close to the temperature probe, adjust the wire or hose so that the distance is between 25 mm and 100 mm, ideally 30 mm. Check the airway sensor on the temperature probe for faults.