2500A Series Operation Manual, 34172 Revision C, March 2008
Specifications &
Performance Verification
4.1 Models
4.2 Specifications
All specifications apply over a 0°C to +55°C range after 30 minutes of warm-up time, unless otherwise
stated.
NOTE: Giga-tronics recommends a calibration cycle of two years for the 2500A.
4.2.1 Frequency
Accuracy: Same As Timebase (after 30 day warm-up)
Resolution: 0.001 Hz
Internal References:
10 MHz
Aging Rate: < 5 X 10
-10
/day
Temperature Stability: < +
2.5 x 10
-8
/
o
C
Table 4-1: Models
1
1. Option 18, Deletes 100 kHz to 2 GHz frequency range
Models Frequency Range
2508A/2508AS 100 kHz to 8 GHz
2520B/2520AS 100 kHz to 20.199 GHz
2526B/2526AS 100 kHz to 26.5 GHz
2540B/2540AS 100 kHz to 40 GHz
4