EasyManua.ls Logo

Giga-tronics 2508A - Appendix A: a Ccessories & Options; Appendix B: Error Messages; Start-Up Error Messages

Default Icon
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2500A Series Operation Manual, 34172 Revision C, March 2008
Error Messages
This appendix lists error and other messages that might be encountered during operation of the instru-
ment. In some cases, encountered errors will be can be remedied by the user, while in other cases, you
might need to contact Giga-tronics support. This appendix consists of the following sections:
“Start-Up Error Messages” - This section lists the messages that might be encountered during the
instrument’s power-up sequence.
“NVRAM Messages” - This section lists messages that might be encountered if the system detects
problems with internal non-volatile memory (NVRAM).
“Remote Error Messages” - This section lists the messages that might be encountered during remote
operation of the instrument.
B.1 Start-Up Error Messages
If the system encounters any problems during the start-up sequence, a message is displayed after start-up
is complete. Typically, you should contact Giga-tronics customer support if any start-up error messages
are encountered. The message that is displayed has the following format:
Error code: xxxxxxxxxxxxxxxxxxxx
Startup Failure, see manual
Where xxxxxxxxxxxxxxxxxxxx is a 20-digit binary number representing the errors that have occurred. The
rightmost digit corresponds to Error #1 in Table B-1, below, and the leftmost digit corresponds to Error
#20. A “1” in a particular bit position indicates an error. For example, the following message displayed
after start-up:
Error code: 00000000000000000111
Startup Failure, see manual
Indicates that Error #1, 2, and 3 shown in Table B-1 have been detected.
1 TIMEBASE_SET_ERROR
2 NVRAM_BATT_FAIL
3 CPU_FPGA_LOAD_FAIL
B

Table of Contents

Related product manuals