EasyManua.ls Logo

Gilson ASPEC - General Flow Guidelines

Gilson ASPEC
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
ASPEC® POSITIVE PRESSURE MANIFOLD | USER'S GUIDE
OPERATION
General Flow Guidelines
The following table provides general process and flow guidelines.
Table 1
SPE Process and Control Guidelines for SPE Steps
SPE Steps
Process
Control
Setting
SPE Control Setting
NOTE
For all low and high flow recommendations, refer to the
SPEFormat Flow Guidelines table below.
Condition SPE High Flow, and then switch to LowFlow
Load SPE Low Flow, based on viscosity
Wash SPE High Flow, based on viscosity
Dry Dry O
Elute SPE Low Flow, based on viscosity
Table 2
SPE Format Flow Guidelines
SPE Format
High Flow
(drops per second)
Low Flow
(drops per second)
6 mL 3 to 5 1
3 mL 1 1 drop per 2 to 3 seconds
1 mL
microplate
0.5 to 1 1 drop per 3 to 4 seconds