6
CONFIGURATION & FIELD CALIBRATION
1.4 CALIBRATION DUE DATE
The CAL DUE date can be set by the workshop and is set
to 30 days by default. The date can be set from 1 to 400
days.
The CAL DUE is reset when the instrument is successfully
calibrated.
The CAL DUE feature has five (5) configurable options:
1. Cal Due Date message is not displayed.
2. Cal Due Date and overdue date messages are
displayed.
Calibration due date is displayed, as shown in Fig.
1.6.
Fig. 1.6 Calibration Due Date
The screen, shown in Fig. 1.7, is displayed when
the Calibration date has expired. i.e. overdue.
Fig. 1.7 Calibration Overdue