32
CONFIGURATION & FIELD CALIBRATION
To accept the highlighted option:
Press and hold OK and the display will move to
the next step.
2.5.1 Update
The calibration data, time and date will now be stored in
the instrument memory. If the calibration period option is in
use, the calibration due date will now be updated
automatically from the calibration period option in the
configuration of the instrument.
2.5.2 Save Only
The new gain setting will be updated but the calibration
due date will not update if the calibration period is in use
from the setup menu. This option may be selected if you
were unable to calibrate all the sensors in the instrument
but wish to save the gain setting for the sensors calibrated
successfully.
2.5.3 Abandon
Neither the gains or the calibration due date will be updated
if this option is highlighted. The instrument will go directly
to the MENU display.