55
SET DEFAULTS
Immediately following the GT Series screen display (Fig.
4.3) and while the instrument identification screen (Fig.
4.4) is displayed,
Press the following buttons in sequence:
LH (INVERT) button
RH (PUMP) button
LH (INVERT) button
RH (PUMP) button
The instrument then begins its warm-up routine, which
lasts approximately 30 seconds. During warm-up, a
countdown timer appears in the top (RH) corner of the
display.
(Refer to Chapter 1 for complete warm-up sequence).
If the correct button sequence is accepted, the countdown
timer alternates with M (menu) symbol.
The instrument now proceeds through the normal start
up sequence, ending with a check of all sensors (only
displayed if a sensor is faulty). If a sensor is faulty, the
orange LED will be active and an audible beep will be
heard.
The instrument displays the configuration & field
calibration menu screen. The alphanumeric screen now
acts as a prompt to help you through the configuration
process.