26 GMC-I Messtechnik GmbH
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006 class B
Interference immunity EN 61326-1:2006
Ambient Conditions
Accuracy
temperature range 0 ... +40C
Operating temperature -10 ... +50 C
Storage temp. range -25 ... +70 C (without batteries)
Relative humidity Up to 75%
(max. 85% during storage/transport),
no condensation allowed
Elevation Max. 2000 m
Calibration interval 1 year (recommended)
Displays
Digital display With additional bar graph or pointer
depending on selection with a.disp
parameter, backlit (transflective);
leading zeros can be suppressed at the
digital display depending on 0.disp
parameter;
overranging indicated with 0L at display;
dimensions: 65 x 36 mm
Limit LED LED lights up red to indicate an exceeded
limit value
LED lights up green to indicate adherence
to the limit value
LED LED lights up red to indicate the presence
of an external voltage (with device
switched off)
or a residual voltage after insulation testing
(U > 50 V)
at LCD Detection of external voltage at the LCD
with the device switched on where U DC >
50 V and U AC > 40 V (50 Hz) for all mea-
suring functions
!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gossen MetraWatt METRISO G1000 and is the answer not in the manual?

Gossen MetraWatt METRISO G1000 Specifications

General IconGeneral
BrandGossen MetraWatt
ModelMETRISO G1000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals