3. ON / OFF
Press on the key. The display switches on and performs the
following sequence:
1. All segments and symbols switch on for a second on the LCD display
to verify that they are functioning properly.
2. It then shows the unit’s firmware version code during one second.
3. Finally, for one second it displays the scale’s maximum capacity and
the interval.
After the power-up sequence the scale is ready for use.
To switch off the unit, press on the same power-on key and keep it pressed
for 2 seconds. An Off message will appear to warn that the device will be
switched off when the key is released.
4. INITIAL RESET TO ZERO
When switching on the unit, it will automatically set to zero. This will be
indicated in the display with the message -00-.
Automatically resetting to zero at the start requires the scale to remain
stable for at least 5 seconds. For as long as the scale is moving, the -00-
indication will be maintained for a maximum of 10 seconds.
If this time period is exceeded without obtaining a stable reading, the display
will show an Err 0 message followed by the weight on the scale.
If the weight on the load receptor is higher than 10% of the scale’s maximum
capacity, the error message Err 0 will be displayed, followed by the weight
on the scale.
The InI-0 option in the configuration menu activates or deactivates the
initial zeroing.