Possible values are 1 (moving average of 2 readings), 2 (4 readings),
3 (8 readings), or 4 (16 readings).
A smaller filter value implies that rapid oscillations in the weight become
more visible and vice versa.
17. WEIGHING OF LIVE ANIMALS
An additional filtering is available for weighing live animals or persons which,
when moving on the load receptor, cause disruption in the weight reading by
exerting a variable force on the load cells.
By activating this filtering in conjunction with the digital filter, an average
reading is obtained that discards the extreme values caused by the shifting
on the load receptor platform.
18. MOVEMENT FILTER
When activating the movement filter, the indicator does not show
inconsistent changes of more than one division with a duration of less than
100ms.
The result is that the indicator retains the last stable value while the weight
has no consistent movement (in the same direction).
Put another way, a sudden quick blow on the scale does not have any effect
on the reading. Although the weight value shown on the display is not
affected, the stability reading in the top left corner of the display will go off
to indicate this situation.
19. INFORMATION TO BE VIEWED
Allows you to select the value that should be shown in the LCD display. The
options are the following:
The indicator shows the weight. It is the default option.
NOT-F
DISP
wEIgh
LIVES