39
Measurement unit: g, kg, oz, lb.
Maximum capacity of the scale. Enter the value, including
the decimal digits.
Division: The smallest increment that the instrument can
measure. Possible values are 1, 2, 5, 10, 20 or 50.
Position of decimal point.
Configuration menu for the instrument’s options
associated with “zero”.
Initial zero setting at start Yes / No.
Allows you to select the limit of the zero-setting device.
Possible options are MAX (zero is allowed for any weight
on the scale) or OIML (follows established rules by the
OIML R76 / EN 45501 technical regulation).
Zero tracking device activated or deactivated.
Show zero indicator in the display Yes/No.
Settings menu for the instrument.
15.1. Settings menu
The settings menu can be directly accessed when switching on the unit. To do
this, switch on the unit and, while the LCD test appears with all segments on,
at the same time press on the and keys (a short press, not
sustained).
CAL
unIt
NAx
dIv
dEC
2ErO
CAL
InI-0
NAx-0
0-trA
0-dIS