Settings and Measurement
3-23
uu
uu
u OTHR Settings Window
Other miscellaneous settings are made here.
Grid ..................................... The Grid parameter for display can be specified as either On or Off.
On: The waveforms are displayed on a grid.
Off. Only the waveforms (without any grid) are displayed.
Screen Saver ...................... Turns off the display if the GL400 is not operated within the specified
interval, thus prolonging the product life of the display.
Off: The screen saver does not operate.
Time: The interval can be specified as 1, 2, 5, 10, 30, or 60 minutes.
View in large print ............... This sets whether or not to switch automatically to the next group of
digital display channels.
Auto Next Screen: The display changes automatically to the next
group of channels.
Fixed: The display can only be changed manually
Power On Start ................... Initiates measurement as soon as the GL400 is turned on.
Disable: Disables the Power On Start function.
Enable: Enables the Power On Start function.
Room Temp. Compensation
.................................. This parameter enables room temperature compensation settings
when thermocouples are used. You can select from either Internal or
External room temperature compensation.
Internal: The GL400's room temperature compensation settings are
used.
External: Select this parameter when measuring compensation other
than that of the GL400.
Note: The Internal setting is normally used.
Setting Selections available Setting method
Grid On, Off ENTER→Select→ENTER
Screen Saver Off, 1, 2, 5, 10, 30, 60 min ENTER→Select→ENTER
View in large print Auto Next Screen, Fixed ENTER→Select→ENTER
Power On Start Enable, Disable ENTER→Select→ENTER
Room Temp. Compensation Internal, External ENTER→Select→ENTER
External Filter Off, On ENTER→Select→ENTER
Burn Out Off, On ENTER→Select→ENTER
Date/Time Date, time settings ENTER→Select→Confirm
Language Japanese, English (US), English (UK) ENTER→Select→ENTER
Return to default settings
Yes, No ENTER→Select→ENTER