Configuration & Commissioning
32
Oxygen Analyzer
The route for setting of back-flushing and auto calibration is Tools setting\Services
(password level 1: 9490)\Back-flushing setup.
Back-flush period is used to set the interval between two back-flashings. The default
setting for auto back-flushing interval is 6 hours (360 minutes). There will be no back-
flushing if the Back-flush period is set to 0.
Back-flush duration is per default set to 30 seconds.
Auto calibration is used to set the interval between two auto calibrations. By default,
the analyzer is set for auto calibration every 24 hours (1440 minutes). There will be no
auto calibration if the Auto Calibration is set to 0.
Figure 7-4: Setting of back-flushing
Freeze output is the freezing time of output signals after calibration or back-flushing.
Default setup is 15 seconds. This parameter can be set either here or in the setting of the
Freeze output under Analog in/out setup. See chapter 7.
+
/
--
+
/
--
+
/
--
+
/
--
<