EasyManua.ls Logo

GREEN INSTRUMENTS G36 - Gas Connection

GREEN INSTRUMENTS G36
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Routine Maintenance
38
9.4 Gas connection
The calibration and sample gas connections to the sensor should always be checked and
ensured no leaking. A leaking connection might result in inaccurate measuring results.
Besides in order to have accurate measuring results, the gasses supplied to the sensor
must have a constant flow and pressure. The pressure and flow rate of the sample gas
should be the same as the pressure and flow rate of the calibration gasses.

Table of Contents