G36
Oxygen Analyzer
Version 2.11 - March 2022 45
For span calibration:
Simulate a current of 20 mA to Analog Input 1. For passive input signal, connect to
terminal INPUT connection 1&3. For active input signal, connect to terminal INPUT
connection 1 & 8.
Select SpanCalCnt and adjust the value in SpanCalCnt until value of analog input
Ai01 in the Main display equal to the MaxDisp value. Save the new value of Span-
CalCnt.
For the calibration of Analog Input 2 (Ai02), select Analog Input 2 in the Analog settings
menu and connect the simulator to Analog Input 2, terminal INPUT connection 2 & 3
(passive input signal) or connection 2 & 8 (active input signal). Then, follow the same
procedure for setting of Analog Input 1.
Open Loop Monitor
No setting is required for this parameter. The default setting is OFF and shall not be
changed.