29
Repeat the calibration by pressing
K at the CAL OK? prompt.
13. Once the calibration is accepted in step5, press
at the ENTER=SAVE prompt and again at
the
ENTER=EXIT prompt to save the new calibration and exit the calibration mode.
- or -
To exit the calibration mode without
saving the new calibration, press K at the ENTER=SAVE
prompt. Then press
at the ENTER=UNDO prompt and again at the ENTER=EXIT
prompt to exit the calibration mode.
Multi - Scale Calibration
When more than one scale is enabled, the prompt Keyin Scl# appears before accessing the calibration
method selections. Key in the scale number to be calibrated and press
. Proceed with a calibration
method as described in Calibration Methods on page 23. After completing a calibration, the
Keyin
Scl#
appears once again. Enter the next scale number to be calibrated, or press K to exit the calibration
mode and save the new calibration data.