REMOTE CONTROL
101
sens:aver:tcon Selects digital filter type.
Parameter: mov (moving), rep (repeating)
Example: sens:aver:tcon_mov (moving digital filter)
sens:aver:tcon? Returns digital filter type.
Parameter: MOV (moving), REP (repeating)
sens:aver:coun Sets digital filter count.
Parameter: 2 ~ 100
Example: sens:aver:coun_100 (filter count 100)
sens:aver:coun? Returns current digital filter count.
Parameter: 2 ~ 100
sens:aver:stat Turns digital filter On/Off.
Parameter: Boolean
Example: sens:aver:stat_1 (digital filter On)
sens:aver:stat? Returns digital filter status, On or Off.
Parameter: Boolean
UNIT command
unit:temp Selects temperature unit, celsius or fahrenheit.
Parameter: c (celsius), f (fahrenheit)
Example: unit:temp_c (temperature unit celsius)
unit:temp? Returns temperature unit, celsius or fahrenheit.
Parameter: C (celsius), F (fahrenheit)
CALCulate command
calc:func Activates advanced measurement functions.
Parameter: rel (relative), max (Max), hold (Hold), dbm
(dBm), db(switches between dB, dB+dBV, and dB+dBm),
math (Math), comp (Compare)
Example: calc:func_math (activate math function)
Example: calc:func_db (activate dB)
calc:func_db (second issue activate dB+dBV(dBm))
calc:func_db (third issue activate dB+dBm(dBV))
calc:func? Returns current advanced measurement functions.
Parameter: rel (relative), max (Max), hold (Hold), dbm
(dBm), DB-V (dB-dBV), DB-M (dB-dBm), math (Math), comp
(Compare)
calc:stat Turns math function On/Off.
Parameter: Boolean
Example: calc:stat_1 (math function On)
calc:stat? Returns math function status, On or Off.
Parameter: Boolean