PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
50
Now, switch to SAVE by using the knob and press [ENTER] to leave the
mode with storing. Also can press [SHIFT][ .] to leave the mode without
storing.
Caution. When proceed the current ad
j
ustment, please
use a Shunt with sufficient Watt to overcome the
problem of temperature rising. A Shunt with
g
eneral
material of Man
g
anin can be used as lon
g
as its
temperature is less than 85.
[Step 6]
Set to [O.C.P.] by using the knob and press [ENTER] getting into Over
Current Protection auto-calibration procedure. Now, the output is in CC
mode (output terminal short).
Switch to SAVE by using the knob and press [ENTER] to leave the mode
with storing. Also can press [SHIFT][ .] to leave the mode without
storing.
PSM-SERIES PROGRAMMABLE POWER SUPPLY
USER MANUAL
51
[Step 7]
Set to [SAVE] and press [ENTER] to complete the calibration procedure
with storing. If want to cancel the calibration, using the knob to set the
cursor to [EXIT] and press [ENTER] to terminate the procedure without
storing.
6-5. Cleaning
To clean the power supply, use a soft cloth dampened in a solution of mild
detergent and water. Do not spray cleaner directly onto the instrument,
since it may leak into the cabinet and cause damage. Do not use chemicals
containing benzine, benzene, toluene, xylene, acetone, or similar solvents.
Do not use abrasive cleaners on any portion of the instrument.